WARNING
The accuracy of SpO2 measurements can be affected by any
of the following:
•
elevated levels of total bilirubin
•
elevated levels of Methemoglobin (MetHb)
•
elevated levels of Carboxyhemoglobin (COHb)
•
hemoglobin synthesis disorders
•
low perfusion at the monitored site
•
the presence of concentrations of some intravascular dyes, sufficient to
change the patient's usual arterial pigmentation
•
patient movement
•
patient conditions such as shivering and smoke inhalation
•
motion artifact
•
painted nails
•
poor oxygen perfusion
•
hypotension or hypertension
•
severe vasoconstriction
•
shock or cardiac arrest
•
venous pulsations or sudden and significant changes in pulse rate
•
proximity to an MRI environment
•
moisture in the sensor
•
excessive ambient light, especially fluorescent
•
the use of the wrong sensor
•
a sensor misapplied or dislodged
•
severe anemia
•
venous congestion
1. Verify that the sensor cable is connected to the monitor.
WARNING
Patient injury risk. The sensor and extension
cable are intended only for connection to pulse co-oximetry
equipment. Do not attempt to connect these cables to a PC
or any similar device. Always follow the sensor
manufacturer's directions for care and use of the sensor.
2. Clean the application site. Remove anything, such as nail polish, that could interfere
with sensor operation.
Note
Do not use disposable sensors on patients who have allergic
reactions to the adhesive.
3. Attach the sensor to the patient according to the manufacturer's directions for use,
observing all warnings and cautions.
WARNING
Patient injury and inaccurate measurement risk.
Do not use tape to secure the sensor to the patient. This can
restrict blood flow, damage the patient's skin, cause
inaccurate readings, and damage the sensor.
168 Patient monitoring
Welch Allyn Connex
®
Devices
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......