The Profile vertical tab appears.
2. Select the desired profile.
The profile change immediately takes effect.
Note
For all of these non-continuous profile changes, saved
episodic patient measurements remain on the Review tab,
and current measurements remain onscreen. Patient and/or
location context, if established, remain onscreen for all non-
continuous profile changes.
Change to the Continuous Monitoring profile
Automatic profile change
If "Allow profile change" has been enabled in Advanced settings, the device attempts to
change to the Continuous Monitoring profile automatically when a continuous sensor
(CO2, RRa, EarlySense) is attached to the device. When the profile change occurs, an
information message appears in the Device Status area to report that change.
Note
The profile change dialogs described in the following Manual profile change
section appear onscreen as appropriate based on established patient
context and/or patient measurements on the device.
Note
An automatic change to the Continuous Monitoring profile occurs only once
for each continuous sensor attachment. If the attempted profile change
fails, or if you switch to an episodic profile after an automatic change to the
Continuous Monitoring profile, you must disconnect and then reconnect a
continuous sensor for another automatic change to occur.
Note
If the device cannot change to the Continuous Monitoring profile, a dialog
indicating that this profile is unavailable in the current configuration
appears.
Manual profile change
To manually change to the Continuous Monitoring profile from any other profile,
complete these steps.
1. Touch the selected profile indicator in the Device Status area.
The Profile vertical tab appears. Based on the boldface text presented, select and
follow the step and substeps below that apply to your patient and the device.
2.
When no patient and/or location context are established and no patient
measurements have been taken or saved,
select the desired profile.
The profile change immediately takes effect.
3.
When patient and/or location context are established but no patient
measurements have been taken or saved:
a. Select the desired profile.
Directions for use
Profiles 87
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......