Change the battery (Connex VSM)
Before removing the battery, shut down the monitor.
1. Turn the monitor upside down to access the battery cover.
2. Locate the battery cover, indicated by
.
3. Insert a coin into the slot and push to open. Choose a coin that fits comfortably in
the slot.
4. Pull the battery out by pulling the battery label, which is visible when you open the
battery cover.
5. Slide in the new battery. Ensure that you insert the new battery in the same
orientation as the old battery.
6. Replace the battery cover by positioning the end below
and then pressing
firmly on the opposite end.
Directions for use
Maintenance and service 181
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......