CAUTION
United States Federal law restricts this monitor to sale,
distribution, or use by or on the order of a physician or licensed healthcare
professional.
CAUTION
Electromagnetic interference risk. The monitor complies with
applicable domestic and international standards for electromagnetic
interference. These standards are intended to minimize medical equipment
electromagnetic interference. Although this monitor is not expected to
present problems to other compliant equipment or be affected by other
compliant devices, interference issues still may occur. As a precaution,
avoid using the monitor in close proximity to other equipment. In the event
that equipment interference is observed, relocate the equipment as
necessary or consult manufacturer's directions for use.
CAUTION
Use only a Class I (grounded) AC power supply cord for
powering this monitor.
CAUTION
Do not use a long press of to power down the monitor when
it is functioning normally. You will lose patient data and configuration
settings.
CAUTION
Never move the monitor or mobile stand by pulling on any of
the cords as this may cause the monitor to tip over or may damage the
cord. Never pull on the power cord when removing it from the power
outlet. When disconnecting the power cord, always grasp the attachment
plug and not the cord. Keep the cord away from liquids, heat, and sharp
edges. Replace the power cord if the strain relief or cord insulation is
damaged or begins to separate from the attachment plug.
CAUTION
Use only the Welch Allyn USB client cable to connect a laptop
computer to the USB client port. Any laptop connected to the monitor must
be running on a battery, a 60601-1-compliant power supply, or a 60601-1-
compliant isolation transformer. While monitoring a patient, you can only
charge the laptop battery if it is connected to 60601-1-compliant, isolated
AC power.
CAUTION
If the touchscreen is not responding properly, refer to the
troubleshooting section. If the problem cannot be resolved, discontinue use
of the monitor and contact an authorized Welch Allyn service center or
qualified service personnel.
CAUTION
Verify patient identity on the monitor after manual or barcode
entry and before printing or transferring patient records.
CAUTION
Keep the monitor outside of MRI suites and any areas marked
for high magnetic or electric field strength.
CAUTION
Do not use the Suretemp to take or monitor the patient’s
temperature during defibrillation or electrosurgery. This may damage the
temperature probe.
CAUTION
Before weighing a patient on any weight scale connected to the
monitor, disconnect all sensors from the patient. Doing so ensures an
accurate weight measurement and significantly reduces electrostatic
discharges that might disrupt the monitor.
20 About warnings and cautions
Welch Allyn Connex
®
Devices
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......