a. Remove continuous sensors from your patient.
b. On the Settings tab, touch the
Monitor
tab.
c. Touch
End monitoring
.
An "End monitoring" dialog appears with the following options:
•
New patient – Clears patient and location context and review data
•
Power down – Clears patient and location context and review data, and
powers down the device
•
Cancel – Retains patient data and returns to Home tab
Note
Ending monitoring deletes all patient measurement data
from the device.
d. Select the desired option and touch
OK
.
The Home tab appears.
e. Repeat step 1 and select the desired profile on the Profile vertical tab.
The profile change immediately takes effect.
4.
When patient and/or location context are established, but no episodic or
continuous trend data has been stored:
a. Select the desired profile.
A confirmation dialog appears asking you to confirm the patient and/or location
context information or select a new patient. For example, if both patient and
location context have been established, the following options appear:
•
Same patient, same location
•
Same patient, different location
•
New patient
b. Select the desired option and touch
OK
.
The profile change immediately takes effect.
Note
If the device was connected to a central station while the Continuous
Monitoring profile was active, changing to any other profile disconnects the
device from the central station.
Directions for use
Profiles 89
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......