Integrated Wall System warnings and cautions
In addition to the preceding warnings and cautions, consider the following when using
the Integrated Wall System.
WARNING
Liquids can damage electronics inside the Connex IWS.
Prevent liquids from spilling on the wall system.
If liquids are spilled on the wall system:
1. Power down the wall system.
2. Disconnect the power plug.
3. Remove the wall system from the wall.
4. Remove battery pack from the wall system.
5. Dry off excess liquid from the wall system.
Note
If liquids possibly entered the wall system, remove the wall
system from use until it has been properly dried, inspected,
and tested by qualified service personnel.
6. Reinstall battery pack.
7. Mount the wall system on the wall.
8. Power on the wall system and verify that it functions normally before using
it.
WARNING
The physical assessment instruments (handles) are designed
for intermittent use. On-time should not exceed 2 minutes. Allow at least
10 minutes off-time between patients.
CAUTION
Welch Allyn is not responsible for the integrity of any wall
mounting interface. Welch Allyn recommends that you contact your
Biomedical Engineering Department or maintenance service to ensure
professional installation, safety, and reliability of any mounting accessory.
CAUTION
Do not use the Suretemp to take or monitor the patient’s
temperature during defibrillation or electrosurgery. This may damage the
temperature probe.
Directions for use
About warnings and cautions 21
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......