d. Touch
OK
.
The General tab will appear.
2. Touch the
Parameters
tab.
3. Touch the
Temperature
tab.
4. Specify settings.
Setting
Action/Description
Unit of measure
Select primary units of measure for the temperature
display on the Home tab
Display temperature conversion
Select to display primary units of measure and secondary
units of measure for the temperature display on the
Home tab.
Default SureTemp Plus site
Select the default site for SureTemp measurements. The
default site applies when clinicians power up the
monitor and each time clinicians remove the temperature
probe from the well.
Select
Last site
to set the default to the site selected
for the last measurement.
Anti-theft return timeout
Select "Disabled" or a timeout value in hours for the
Braun 6000 to lock after being removed from the dock.
Mode
Select
None
,
Technique Compensation
, or
Unadjusted
as the Braun 6000 compensation mode.
Technique Compensation improves reading accuracy by
detecting placement of the probe in the ear canal.
Unadjusted places thermometer in mode to detect raw
ear temperature only.
Enable pulse timer
Select to enable pulse timer on the Braun 6000 handle.
Enable Celsius only selection
Select to enable Celsius only mode, thus disabling the C/
F button on the Braun 6000 handle as well as the
hardware switch.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch
Exit
.
Specify the manual parameters
The Manual Parameters frame is in the lower right corner of the Home tab. You can
manually enter values for parameters in the frame. In Advanced settings, you can specify
which parameters appear in the frame and enable manual overrides of other parameter
measurements displayed on the device.
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter the
Advanced settings code
.
d. Touch
OK
.
The General tab appears.
2. Touch the
Parameters
tab.
3. Touch the
Manual
tab.
198 Advanced settings
Welch Allyn Connex
®
Devices
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......