The General tab appears.
2. Touch the
Parameters
tab.
3. Touch the
Pulse rate
tab.
4. Specify settings.
Setting
Action/Description
Display source
Select this to show the source of pulse rate
measurements (NIBP or SpO2) on the Home tab.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch
Exit
.
Specify advanced etCO2 settings
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter the
Advanced settings code
.
d. Touch
OK
.
The General tab appears.
2. Touch the
Parameters
tab.
3. Touch the
etCO2
tab.
4. Specify settings.
Setting
Action/Description
Default view
Select a numeric view or a waveform view as the
primary etCO2 display on the Home tab.
Display FiCO2
Select to display FiCO2 on the Home tab.
Waveform scale default
Select the default waveform scale in etCO2
measurements.
Unit of measure
Select primary units of measure for the etCO2 display on
the Home tab.
Check calibration
Select to start CO2 calibration check.
Calibrate
Select to start CO2 calibration.
Sweep speed default
Select the default waveform sweep speed for the CO2
waveform view on the Home tab.
Enable sampling line dialog
Select to enable the sampling line dialog to be displayed
when connecting a sampling line to the device.
BTPS compensation
Select to enable automatic adjustments for BTPS (body
temperature pressure saturated) to improve the accuracy
of CO2 measurements.
Calibration due hours
Displays time remaining until CO2 sensor calibration is
required.
Maintenance due hours
Displays time remaining until CO2 periodic sensor
maintenance is required.
Last calibration
Displays date (XX/XX/XXXX) and time (00:00:00) of last
calibration.
Annual calibration
Displays anniversary date (XX/XX) of annual calibration.
194 Advanced settings
Welch Allyn Connex
®
Devices
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......