CO2 specifications (for additional information, refer to the directions for use that accompanies
each accessory)
Dispose of sampled gases according to facility requirements
or local regulations.
Sampling rate
20 samples per second
Calibration interval
Initial: After 1200 operating hours
Subsequent: After 4000 operating hours or annually
(whichever comes first)
Periodic service
After 30,000 operating hours
etCO2
Units of measure
mmHg, kPa; user-selectable
Display range
0 to 150 mmHg (0.0 to 20.0 kPa)
Resolution
1 mmHg, 0.1 kPa
Frequency response
The device maintains etCO2 accuracy up to 80 breaths/min.
From 81 to 150 bpm, accuracy is ±12%.
FiCO2
Units of measure
mmHg, kPa; user-selectable
Display range
0 to 150 mmHg (0.0 to 20.0 kPa)
Resolution
1 mmHg, 0.1 kPa
IPI
Display range
1 to 10
RR
Unit of measure
breaths per minute
Display range
0 to 150 breaths per minute
Accuracy
0 to 70 breaths per minute: ±1 breath per minute
71 to 120 breaths per minute: ±2 breaths per minute
121 to 150 breaths per minute: ±3 breaths per minute
Resolution
1 breath per minute
1
For breath rates above 80 bpm, accuracy is 4 mmHg or ±12% of reading (whichever is greater) for etCO2 values
exceeding 18 mmHg when measured in accordance with ISO 80601-2-55.
Directions for use
Specifications 231
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......