Control Board
5-8
Force EZ-C Series Service Manual
Power Supply Supervisor Circuit
The power supply supervisor circuit (U14), a MAX691, generates a Reset signal
and a Reset\ signal for the main microcontroller (U5) if the power supply voltage
to the Control board drops below 4.65 V. Reset\ also places the PSD835G2 (U3)
and the PSD835G2 (U6) in sleep mode and disables the 2K x 8 external static
RAM.
A/D and D/A Conversion
Each 80C562 microcontroller (U5 and U11) contains an 8-channel multiplexed 8-
bit A/D converter. Incorporating gain scaling relays in the sense circuits on the
Power Supply/RF board and prescaling based on the expected input voltage or
current values enhances the resolution of voltage and current sense inputs.
The main microcontroller senses redundant RF output current and voltage from
additional sense circuits located on the Power Supply/RF board. The system does
not gain scale this information since it is for dosage monitoring only.
An MP7226 quad digital-to-analog (D/A) converter (U15) provides 4-channel 8-
bit D/A capabilities for the feedback microcontroller to output 0 to 5 Vdc analog
voltages.
Waveform Generation (T_ON ASIC)
A dedicated 89C54 microcontroller (U9) generates the RF waveforms (T_ON\)
for the RF output amplifier on the Power Supply/RF board. The microcontroller
functions as an application-specific integrated circuit (ASIC) performing an
endless series of repetitive tasks while enabled.
The feedback microcontroller (U11) holds the T_ON ASIC (U9) in a reset state
until the feedback microcontroller detects a valid activation request. After
validating the request, the feedback microcontroller releases the T_ON ASIC
from reset and communicates a 4-bit code that represents the generator mode to
be activated. Six activation codes are acceptable:
•
0: bipolar
•
2: pure, low 2, and low 3
•
3: blend
•
7: low 1
•
8: high 1
•
C: high 2
The Force EZ-C Series Electrosurgical Generator does not use Codes 1, 4, 5, 6,
9,A,B, and D-F.
Each code generates a unique waveform pattern to be delivered to the RF output
stage of the generator. The T_ON ASIC reads and evaluates the code and, if the
code value is acceptable, repetitively generates the appropriate waveform until
the activation request ends. After the request ends, the feedback microcontroller
places the T_ON ASIC back into reset.
Содержание Force EZ - C Series
Страница 1: ...Service Manual Valleylab Force EZ C Series Electrosurgical Generator with Instant Response Technology...
Страница 6: ...vi Force EZ C Series Service Manual...
Страница 20: ...1 6 Force EZ C Service Manual...
Страница 64: ...4 30 Force EZ C Series Service Manual...
Страница 156: ...7 26 Force EZ C Series Service Manual...
Страница 158: ...Interconnect Diagram 8 2 Force EZ C Series Service Manual Interconnect Diagram Figure 8 1 Electric cable connections...
Страница 208: ...10 18 Force EZ C Series Service Manual...