Connections for Monopolar Surgery
Setup, T
e
st
s, and Adjustment
s
Force EZ-C Series Service Manual
6-5
Figure 6-3.
Bipolar connection
(handswitching instrument)
Setting the Bipolar Output
1.
To increase (+) the power, turn the Bipolar Power Control knob clockwise.
To decrease (-) the power, turn the knob counterclockwise. The maximum
power setting for bipolar output is 70 watts.
2.
To display and use the previous power setting, press the Pure and Blend
buttons simultaneously.
Connections for Monopolar Surgery
If you plan to use a footswitching monopolar instrument, you must connect a
Valleylab monopolar footswitch. You may also use a footswitch to activate a
handswitching instrument.
For most procedures, you will connect only one monopolar instrument
(handswitching or footswitching).
Caution
Set power levels to the lowest setting before testing an accessory.
handswitching
instrument
Warning
Electric Shock Hazard —
• Do not connect wet accessories to the generator.
• Ensure that all accessories and adapters are correctly connected and that no
metal is exposed.
Use only a Valleylab monopolar footswitch with the Force EZ-C Series
Electrosurgical Generator. Use of an incompatible footswitch may cause
unexpected output.
The instrument receptacles on this generator accept only one instrument at a
time. Do not attempt to connect more than one instrument at a time into a given
receptacle. Doing so will cause simultaneous activation of the instruments.
Содержание Force EZ - C Series
Страница 1: ...Service Manual Valleylab Force EZ C Series Electrosurgical Generator with Instant Response Technology...
Страница 6: ...vi Force EZ C Series Service Manual...
Страница 20: ...1 6 Force EZ C Service Manual...
Страница 64: ...4 30 Force EZ C Series Service Manual...
Страница 156: ...7 26 Force EZ C Series Service Manual...
Страница 158: ...Interconnect Diagram 8 2 Force EZ C Series Service Manual Interconnect Diagram Figure 8 1 Electric cable connections...
Страница 208: ...10 18 Force EZ C Series Service Manual...