Front Panel Knob
Replacement Procedures
Force EZ-C Series Service Manual
8-11
Important
Front Panel Knob
Equipment
•
Phillips screwdriver
•
Pliers
Procedure
1.
Turn off the generator and disconnect the power cord from the wall receptacle.
Use new knob shaft clips.
2.
Remove the front panel assembly (observe all warnings and cautions). Refer
to
Front Panel
earlier in this section.
3.
Using pliers, pry the clip off each knob shaft. Then, pull the knobs out.
Discard the clips.
4.
For each knob, insert the knob shaft into the front panel and install the new
clip that secures the knob to the front panel.
5.
Reinstall the front panel assembly. Refer to
Front Pane
l earlier in this section.
Warning
Electric Shock Hazard
– To allow stored energy to dissipate after disconnecting
power, wait at least five minutes before replacing parts.
Caution
The generator contains electrostatic-sensitive components. When repairing the
generator, work at a static-control workstation. Wear a grounding strap when
handling electrostatic-sensitive components,
except
when working on an
energized generator. Handle circuit boards by their nonconductive edges. Use an
antistatic container for transport of electrostatic-sensitive components and circuit
boards.
Содержание Force EZ - C Series
Страница 1: ...Service Manual Valleylab Force EZ C Series Electrosurgical Generator with Instant Response Technology...
Страница 6: ...vi Force EZ C Series Service Manual...
Страница 20: ...1 6 Force EZ C Service Manual...
Страница 64: ...4 30 Force EZ C Series Service Manual...
Страница 156: ...7 26 Force EZ C Series Service Manual...
Страница 158: ...Interconnect Diagram 8 2 Force EZ C Series Service Manual Interconnect Diagram Figure 8 1 Electric cable connections...
Страница 208: ...10 18 Force EZ C Series Service Manual...