Correcting Malfunctions
7-6
Force EZ-C Series Service Manual
Rear panel footswitch
will not activate output
1.
Rear panel monopolar footswitch
activates only handswitching
instrument connected to
Monopolar Instrument receptacle
1.
Use a handswitching instrument or connect the
footswitching instrument to a universal
adapter. Then connect the adapter to the
Monopolar Instrument receptacle for
handswitching output.
2.
Faulty connection from Footswitch
board to Power Supply/RF board
2.
Check the ribbon cable and its connections.
3.
Malfunctioning or damaged rear
panel footswitch receptacle
3.
Replace the Footswitch board. Refer to
Footswitch Board
in Section 8.
4.
Footswitch activation signal lost
on Control board
4.
Replace the Control board. Refer to
Control
Board
in Section 8.
5.
Malfunctioning footswitch
interface circuitry on Power
Supply/RF board
5.
Troubleshoot components on the Power
Supply/RF board:
•
Check the DC/DC converter (U11).
•
Check optical isolators ISO1, ISO2, and ISO3.
Situation
Possible Cause
Recommended Action
Содержание Force EZ - C Series
Страница 1: ...Service Manual Valleylab Force EZ C Series Electrosurgical Generator with Instant Response Technology...
Страница 6: ...vi Force EZ C Series Service Manual...
Страница 20: ...1 6 Force EZ C Service Manual...
Страница 64: ...4 30 Force EZ C Series Service Manual...
Страница 156: ...7 26 Force EZ C Series Service Manual...
Страница 158: ...Interconnect Diagram 8 2 Force EZ C Series Service Manual Interconnect Diagram Figure 8 1 Electric cable connections...
Страница 208: ...10 18 Force EZ C Series Service Manual...