Responding to System Alarms
7-12
Force EZ-C Series Service Manual
13
Master microcontroller unable to access
GEN_SEM semaphore.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
14
Internal diagnostics. Calibration data
checksum error on master microcontroller.
Turn the power switch off (O) then on (I) again.
If error reappears, replace the control board.
Refer to
Control Board Replacement
in Section 8.
16
Diagnostics/microcontroller malfunction.
T_ON average test failed.
Calibrate the ECON factor (refer to Section 5 for
instructions).
If the alarm number reappears, refer
Correcting T_ON
ASIC Malfunctions
in this section.
17
18
19
Isns and/or Vsns voltage detected without
activation.
REM circuit failure. REM oscillator
frequency outside acceptable operating
range.
Overvoltage detected on +5V power
supply.
Do not attempt to use the generator. Record the
number and call the Valleylab Service Center.
30
31
32
40
50
51
54
Software malfunction.
audio_state
value
outside state range.
Software malfunction. Invalid alarm
number.
Software malfunction.
alarm_state
value
outside state range.
Software malfunction.
selection_state
[button_num]
value outside state range.
Software malfunction.
encoder_state
[which_encoder]
value outside state
range.
Software malfunction.
which_encoder
value outside state range.
Software malfunction.
user_mode_index
value outside state range.
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call the
Valleylab Service Center.
Number
Description
Recommended Action
Содержание Force EZ - C Series
Страница 1: ...Service Manual Valleylab Force EZ C Series Electrosurgical Generator with Instant Response Technology...
Страница 6: ...vi Force EZ C Series Service Manual...
Страница 20: ...1 6 Force EZ C Service Manual...
Страница 64: ...4 30 Force EZ C Series Service Manual...
Страница 156: ...7 26 Force EZ C Series Service Manual...
Страница 158: ...Interconnect Diagram 8 2 Force EZ C Series Service Manual Interconnect Diagram Figure 8 1 Electric cable connections...
Страница 208: ...10 18 Force EZ C Series Service Manual...