Calibrating the Force EZ-C Series Electrosurgical Generator
Setup, T
e
st
s, and Adjustment
s
Force EZ-C Series Service Manual
6-35
Important
Start with a low Z factor and
increase it only until you obtain
the desired current. Do not
continue to increase it beyond
where you first note the correct
output current.
3.
Adjust the Z factor for the blend mode.
a.
Press the Blend button.
b.
Press the footswitch cut pedal and check the voltmeter for a reading
equivalent to 1.000 +0/–.02 A
rms
(0.98 to 1.00 A
rms
).
c.
Stop activation. If the output current was not within the stated range, use
the Cut and Coag Power Control knobs to adjust the Z factor.
Start with a low value. Adjust until the current is as close as possible to 1.00
A
rms
, but not over 1.00 A
rms
.
Repeat 3b and 3c until the output current is within the stated range.
4.
Disconnect the test cables and remove the 200 ohm resistor from monopolar
output.
5.
To save the reactance gain calibration and go to calibration step 8, turn the
Bipolar Power Control knob clockwise.
Next, go to
Calibration Step 8 – Check and adjust the ECON factor
.
Calibration Step 8 – Check and Adjust the ECON Factor
Equipment
•
Two short test cables (less than 24 inches long) with banana plugs
•
Current transformer (such as the Pearson Model 411 or equivalent)
•
True RMS voltmeter (such as the Fluke 8920 or equivalent)
•
30, 100, and 500 ohm noninductive power resistors (such as the Dale NH-250)
•
Bipolar footswitch and monopolar footswitch
Checking Bipolar Output
1.
Verify that the Bipolar display shows calibration step 8.
2.
Connect the test equipment for bipolar output.
a.
Connect the two test cables to the Bipolar Instrument receptacle.
b.
Connect the 30 ohm resistor across the output jacks at the end of the test
cables.
c.
Connect the bipolar footswitch to the Bipolar Footswitch receptacle on the
rear panel.
or
Connect the monopolar footswitch to the Footswitch receptacle on the front
panel and press the Footswitch Selector button until the left arrow indicator
illuminates green.
Содержание Force EZ - C Series
Страница 1: ...Service Manual Valleylab Force EZ C Series Electrosurgical Generator with Instant Response Technology...
Страница 6: ...vi Force EZ C Series Service Manual...
Страница 20: ...1 6 Force EZ C Service Manual...
Страница 64: ...4 30 Force EZ C Series Service Manual...
Страница 156: ...7 26 Force EZ C Series Service Manual...
Страница 158: ...Interconnect Diagram 8 2 Force EZ C Series Service Manual Interconnect Diagram Figure 8 1 Electric cable connections...
Страница 208: ...10 18 Force EZ C Series Service Manual...