7-24
Force EZ-C Series Service Manual
Important
Correcting Battery-Backed RAM Malfunctions
Equipment:
•
Phillips screwdriver
•
Surface mount, quad pack chip extractor
Procedure
1.
Turn off the generator and disconnect the power cord from the wall receptacle.
2.
Loosen the four screws that secure the cover to the chassis. Lift the cover off
the chassis.
3.
Remove the Control board.
a.
Remove the screw that secures the retention bracket to the left rear heat
sink.
b.
Unlock the connector on the Control board and disconnect the Display
board ribbon cable from the Control board.
c.
Carefully slide the Control board straight up to disconnect it from the
Power Supply/RF board.
Internal memory loses all stored
data (refer to Section 4, Internal
Memory) when you remove the
battery .
4.
Replace the battery.
a.
Grasp the battery and snap it out of the socket.
b.
Position the new 3 V button-cell battery so that the positive side of the
battery is facing out (visible). Refer to
Control board Components
in
Section 9 for the Valleylab replacement battery or call the Valleylab
Service Center for information about approved battery replacements.
c.
Snap the new battery into the socket so it is firmly seated.
Warning
Electric Shock Hazard
– To allow stored energy to dissipate after disconnecting
power, wait at least five minutes before replacing parts.
Electric Shock Hazard
– Do not touch any exposed wiring or conductive surfaces
while the generator is disassembled and energized. Never wear a grounding
strap when working on an energized generator.
Caution
The generator contains electrostatic-sensitive components. When repairing the
generator, work at a static-control workstation. Wear a grounding strap when
handling electrostatic-sensitive components,
except
when working on an
energized generator. Handle circuit boards by their nonconductive edges. Use an
antistatic container for transport of electrostatic-sensitive components and circuit
boards.
Содержание Force EZ - C Series
Страница 1: ...Service Manual Valleylab Force EZ C Series Electrosurgical Generator with Instant Response Technology...
Страница 6: ...vi Force EZ C Series Service Manual...
Страница 20: ...1 6 Force EZ C Service Manual...
Страница 64: ...4 30 Force EZ C Series Service Manual...
Страница 156: ...7 26 Force EZ C Series Service Manual...
Страница 158: ...Interconnect Diagram 8 2 Force EZ C Series Service Manual Interconnect Diagram Figure 8 1 Electric cable connections...
Страница 208: ...10 18 Force EZ C Series Service Manual...