MAN205-0195042EN Rev. A
Aug. 2011
Page 77 of 137 pages
8.3
Switching to the Internal Steam Generator
If the building steam supply fails:
1.
Open the generator electric box, switch ON the 3 Phase circuit breaker with its
Trip Coil, and ensure all the other circuit breakers are ON.
2.
Observe the generator pressure gauge is at working pressure, (4.5 bar).
3.
Turn OFF
Building steam inlet
and turn ON
Generator inlet steam
using the
ON-OFF-ON switch on the front panel of the autoclave.
4.
Check that the Jacket pressure gauge, reaches 2.4 bar
5.
Continue to work as normal.
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...