MAN205-0195042EN Rev. A
Aug. 2011
Page 59 of 137 pages
5.
PRINTER
5.1
Printer Output
The autoclave is equipped with a character printer, which prints a detailed history of
each cycle performed by the instrument (for the record or for subsequent consideration).
The printing is made on plain paper with 24 characters per line and contains the
following information:
Software version
Real time
Selected program
Sterilization pressure
Sterilization temperature
Sterilization time
Summary of performed cycle and identification hints.
When the sterilization cycle begins the printer starts printing the above data.
After the preliminary printing, the autoclave starts performing the sequence of operations
of the cycle. The measured values of temperature and pressure are printed at fixed time
intervals, according to the phase of the process, as shown in the table below.
The data is printed from the bottom up, beginning with the date and ending with ―O.K.‖
for a complete cycle or ―FAIL‖ for an aborted cycle.
For an example of a typical printout, see next page.
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...