MAN205-0195042EN Rev. A
Aug. 2011
Page 15 of 137 pages
Performing Safety Tests
Check the tightness of all the nuts and screws.
Test the earth continuity
– use a Megaohm Tester.
Check current leakage. Check Voltage and Amperage.
Adjust the legs of the transfer carriage to equal the height of loading cart to
rails in the sterilization chamber.
Turn ON the steam generator.
Check connection to steam generator. Open the steam. Manually test the
steam valves by over-riding the appropriate solenoid valve. If there are no
leaks,
–leave the steam manual inlet taps open.
Run a cycle.
Call for an authorized inspector to test and approve the autoclave; generator
and, assembled system.