MAN205-0195042EN Rev. A
Aug. 2011
Page 134 of 137 pages
Fail Downloading Gain and Offset!
Description:
This message may be displayed during download process, when the
program is trying to download Gain and Offset values.
Reason:
Probably a communication failure between PCS Target and the PC, or the
PCS target platform is damaged and cannot save data on flash memory.
Tips:
1. Check if the communication RS232 cable is connected. If it is a modem
connection, check if the modem cable is connected. It may be that the remote
modem cable is disconnected or RS232 cable between the remote modem and
target platform is disconnected.
2. Try to change the target electronic card.
Error Opening Type file:
File Name
Description:
This message may be displayed during the downloading process, when
the program is trying to open the ADMC compatible type file (*.mdb) in order to get
parameters data.
Reason:
The file may be damaged or used by another application.
Tips:
Try to download another file. If operation succeeds, try to find what application is
using the file and in the worst case, restart the PC. If the problem persists,
probably the files are damaged. In this case try to replace the file with an identical
file from the ADMC installation CD under 'Support' directory.
Invalid Actual High value!
Invalid Actual Low value!
Invalid Read High value!
Invalid Read Low value!
Description:
This message is displayed when the user is trying to calculate Gain and
Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
An incorrect number or non-numeric data has been inserted in one of the text
boxes.
Tips:
Type the correct numeric data in the text box.
The Actual High value is not in range!
The Actual Low value is not in range!
The Read High value is not in range!
The Read Low value is not in range!
Description:
This message is displayed when the user is trying to calculate Gain and
Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
An incorrect number has been inserted in one of the text boxes. A correct
number must be entered: 1<Number<1000.
Tips:
Type the right number that is included in the above range.
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...