MAN205-0195042EN Rev. A
Aug. 2011
Page 94 of 137 pages
4.
Slide the new endbell assembly onto the shaft as far as the brush retention clip
will permit.
5.
Index the endbell wires to mark made in step #1. Align endbell to the slots in
the motor housing. Pull the brush retaining clip out while pressing the endbell
flushes to the motor housing.
6.
Connect the black 115V (brown 230V) wire to the pump switch if applicable.
7.
Install the motor assembly screws along with the green/yellow wire (115V
only). Torque to 12º18 m. lbs.
Perform Hi-Pot Test as follows:
1.
The system conductors must be shorted
together, and all switches must be turned
―ON‖.
2.
Hi-pot voltage is applied between the
shorted conductors and accessible dead
metal. The voltage frequency should be
within the range of 40-70 hertz.
3.
Voltage should be applied in accordance
with either condition ―A‖ or condition ―B‖
per the table below.
CONDITION A
CONDITION B
Appliance
Rating
Potential
Volts
Time
Seconds
Potential
Volts
Time
Seconds
Less than 100
600
1
100
– 130 volts
1000
60
1200
1
210
– 240 volts 1480
60
1780
1
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...