MAN205-0195042EN Rev. A
Aug. 2011
Page 108 of 137 pages
2.6
If the safety valve did not blow at the pre-set pressure stop the operation of
the autoclave, decrease the pressure to atmospheric pressure and replace
the safety valve with a new one.
2.7
If the safety valve was replaced, repeat step 2.1
—2.5 to test the new one.
2.8
Close solenoid valve 91 (screw's slot is horizontal).
2.9
Readjust the pressure regulator to the original position. Use the marking as
mentioned above in Step 2.1
Warning!
It is strictly forbidden for any person to enter the service area behind the service
access panels, except the trained technician.
If, for any reason it is necessary to open the service access panel, the person
must shut the system by pressing the emergency switch, withdrawing the key and
keeping the key on him, to prevent accidents and injuries.
The technician must follow the instructions in the SERVICE INSTRUCTIONS,
“Working with power ON in Service Area” Section.