MAN205-0195042EN Rev. A
Aug. 2011
Page 41 of 137 pages
The 12 cycles, names of programs and relevant temperatures are listed below:
Cycle no.
Program
Temp ºC
1, 2, 3
Un-wrapped
134
4, 5, 7
Wrapped
134
6
Wrapped
121
8, 9, 10
Slow Exhaust
121
11
B&D test
134
12
Leak test
—
Second Row:
This row displays the temperature in the chamber, pressure in the chamber, and jacket
pressure.
Third Row:
This row is used to display messages to the operator such as DOOR1 OPEN.
Fourth Row:
On the bottom row, the names of the commanding keys appear. Under the display there
are three keys without symbols or names. These keys receive their commands by the
name that appears above them in the 4th row of the display. During a cycle, the fourth
row is used to display the timers, such as STE Time 02:25, or during pre-vacuum, to
display the stage of the pulse, such as puls1
– pump to 21K.
Second Row
Third Row