MAN205-0195042EN Rev. A
Aug. 2011
Page 52 of 137 pages
4.3.4
Changing Cycle Parameters
The screen that allows for parametric changes appears as follows:
1.
Press CLEAR or ENTER before entering the code to return to the Stand By
screen; the RUN & FAIL LEDs are lit.
2.
Enter the correct code via the UP/DN keys.
3.
Press ENTER to proceed to the next screen.
4.
Press the UP/DN keys to change all parametric values.
Cycle Setup Parameters
CODE 1 OR 11 PARAMETERS
STE Temp
– Required sterilization temperature for the process
Entry code
1
Resolution
1º
Minimum value
105ºC (221 ºF)
Maximum value
137ºC (278 ºF)
STE Time
– Required sterilization time for the process
Entry code
1
Resolution
1 minute
Minimum value
1 minute
Maximum value
99 minutes
Dry Time
– Required drying time for the process
Entry code
1
Resolution
1 minute
Minimum value
Ø minutes
Maximum value
99 minutes
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...