Doors Number
After entering the language, the following message is displayed:
This autoclave has two door, press ØØØ2. Press ENTER, and the system moves to the
next screen.
Door Type
There are 3 different types for autoclaves
– Manual Door, Automatic Hinged Door, or
Automatic Sliding (horizontal or vertical) Door.
In this autoclave, the doors are Vertical sliding doors; the door type is (ØØØ2). Enter
these values either through the UP/DN keys and then press ENTER, or enter the digits
whereby the screen will automatically progress to the next display after entering the
fourth digit.
Atmospheric Pressure
The standard atmospheric pressure at the sea level is 100 kPa and for every 100m
above sea level the atmospheric pressure drops by 1kPa, and for every 100m below sea
level the atmospheric pressure increases by 1kPa. For example, at an altitude of 2500m,
the Atmospheric Pressure will be 75 kPa.
(The standard atmospheric pressure at Sea Level is 14.5psi. For every 1000ft. above
sea level, the atmospheric pressure drops 0.45psi, and for every 1000ft below sea level,
the atmospheric pressure increases 0.45psi.)
When a door of the autoclave is open then set Atm. Pres. The parameter entered into
the program memory will serve as reference for correct pressure control and monitoring.
If pressure is set correctly, then press ENTER and the next screen appears. If pressure
needs to be altered, this may be done either through the digital key or the UP/DN keys.
F
– 0 Mode
This mode enables F
– 0 calculations, the processing being based on data fed by two
temperature sensors. One mounted in chamber and one on the condensate pipeline.
The F - 0 calculations is performed for cycles with sterilization temperatures in the range
of 101ºC
–134ºC (215ºF - 273ºF), especially used in programs 8–10.
F
– 0 = 0 (unable)
F
– 1 = total heating time + total cycle time + cooling time.
F
– 2 = compensate the difference between heating time, cooling time and exhaust.
Calculates using factor 1.5 of the total accumulated heating and cooling time and
converts that into the equivalent sterilization time.
Temp F
– 0 = determines the temperature at which F – 0 calculations should begin.
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...