MAN205-0195042EN Rev. A
Aug. 2011
Page 12 of 137 pages
Drain
Connect the following outlets directly to the drain funnel. or connect them
through a drain collector pipe of 4-
6‖ that should be covered. The drain
system should be vented.
General drain outlet by a ¾" pipe.
Vacuum pump by a 1" pipe
Steam
Connect the building steam supply to the autoclave inlet by a ¾" pipe.
The steam supply pressure should be between 3.5
–4.5 bar (50-65 psi).
(See also Utilities Table).
The steam trap on the building steam supply should not exceed 2 meters
from the machine.