MAN205-0195042EN Rev. A
Aug. 2011
Page 45 of 137 pages
4.3.2
Selecting a Cycle
A cycle can only be selected in Stand By from the Stand By / Ready screen which
displays the cycle number and name in the top row.
To select a cycle:
1.
Press the UP/DN keys to change the cycle where pressing UP moves to the
next program and pressing DOWN moves to the previous program.
2.
Press on a number; press key 4 to select program 4
Selecting a 2 digit cycle:
For programs with double digits (for program 10 or above), both numbers are to be
pressed within 2 seconds of each other. For example, to choose program 12, number 1
must be pressed, and within 2 seconds number 2 must be pressed.
Show Parameters
After selecting the cycle, when pressing the SHOW command key the following screen
is displayed:
The message displayed in the 3rd row describes the following:
The temperature; the sterilization temperature
The time; the sterilization time
D-Time; drying time.
This screen is displayed for 10 seconds after which it automatically returns to the ST.BY
screen.
This command key is not active for cycle 12
– Leak Test.
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...