3.1.3
GICAR Water Level Controller
The water level control system consists of three electrodes. (HIGH and LOW water
level), and a reference electrode, all connected to the
GICAR
2DRLHG controller.
The long electrode which is continuously immersed in the water, acts as a reference and
is connected to ground.
The medium electrode (LOW level electrode) controls the water heaters. When it stops
sensing water, the
GICAR
controller stops the current to the water heaters. When it
senses water, the
GICAR
controller restores current to the water heaters.
The short electrode (HIGH level electrode), controls the water pump. When it senses
water, the
GICAR
controller stops the current to the water pump and simultanausly
closes water supply shut off valve. When it stops s ensing water, the
GICAR
controller
restores the current to the water pump and simultanausly opens water supply shut off
valve.
GICAR 2DRLH
S
R
1
S
M
1
S
m
1
S
R
2
S
M
2
S
m
2
RL1
RL2
S
E
S
E
L N
R
L
1
R
L
2
Water
Pump
Heater
Electrode
Low Low
Reference
Electrode
low
(heater)
Electrode
high
(water )
44
45
46
VAC
1
2
3
Dimensions in mm
Front of Generator
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...