MAN205-0195042EN Rev. A
Aug. 2011
Page 98 of 137 pages
9.7.3
Replacing the Heaters
Disconnect the steam generator from the electrical source. Release the steam pressure
from the autoclave and the steam generator.
1.
Flush the water out of the steam generator then close the drain valve.
2.
Remove the heaters cover.
3.
Remove the wire connections from the heaters.
4.
Unscrew the heaters and remove them from the front plate of the
generator.
5.
Replace new heaters. Ensure that they are tightened well in order to
avoid leakage.
6.
Reconnect the wire connections.
7.
Replace the heaters cover.
8.
Test the steam generator.
Heaters