Replacing the Water Level Electrodes
1.
Turn OFF the electric power and discharge the steam pressure from the
autoclave and the steam generator.
2.
Disconnect the wire from the electrode connections.
Note:
Mark the wires before removing them. It is essential to return the correct wire to
the correct electrode!
3.
Remove the electrode.
4.
Insert a new electrode into their place.
5.
Set the height as per the drawing on the below, and tighten the fixing
nut to avoid leakage.
Note:
The setting of the electrode height is essential the proper working of the
generator!
6.
Reconnect the wire to the electrodes Return the correct wire to the
correct electrode.
7.
Test the unit.
Dimensions in mm
Front of Generator