9.7.4
Replacing the Generator Side Gauge Glass Indicator Tube
Note:
On each side of the glass tube, there are stainless steel balls that act as a safety
device by preventing the escape of steam and boiling water if the glass is
broken. Before replacing the glass water tube, switch OFF the electrical power
and release steam pressure from the vessel.
1. Drain the water.
2. Remove the protective cover from the glass tube.
3. Loosen the two nuts on both sides of the glass tube. Slip the upper one down,
and then take off the tube.
4. Replace the tube with a new one.
5. Replace the two nuts and tighten gently.
Glass tube