MAN205-0195042EN Rev. A
Aug. 2011
Page 96 of 137 pages
9.7.2
Cleaning and Replacing the Water Level Electrodes
The three electrodes of the water level control system are located on top of a tank
mounted on the side of the steam generator.
Cleaning the Water Level Electrodes
1.
Turn OFF the electric power and discharge the steam pressure from the
autoclave and the steam generator.
2.
Disconnect the wire from the electrode connections.
Note:
Mark the wires before removing them. It is essential to return the correct wire to
the correct electrode!
3.
Mark the electrodes before removing them.
4.
Remove the electrode.
5.
Clean the electrode.
6.
Reassemble the electrodes into their places as marked in step 3.
7.
Set the height as per the drawing on the below.
Note:
The setting of the electrode height is essential the proper working of the
generator!
8.
Tighten the nuts to avoid leakage.
9.
Reconnect the wire to the electrodes Return the correct wire to the
correct electrode.
10.
Test the unit.
Dimensions in mm
Front of Generator
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...