MAN205-0195042EN Rev. A
Aug. 2011
Page 129 of 137 pages
Type File Button
This button enables the user to select a target type file (e.g. in the calibration window
below, the file is: "C:\pcs\dbase\LabTn2.DWN"). Pressing this button will display 'Open
file' dialog box. If the program will find a file with the same name in ADMC directory, in
the Inputs table, the Analog inputs names will be displayed in color.
When the program is connected, by double clicking on one of the input table rows, the
program will display the calibration tool. With this tool you can calibrate the selected
analog input.
Description of the Calibration Tool
Gain text box
– use it for manual inserting of the Gain value.
Offset text box
– use it for manual inserting of the Offset value.
Min and Max
– limits for Gain and Offset values. If the ADMC compatible type file is not
found, Min and Max labels will show "N/A".
Note:
When Min and Max are N/A, you must be very careful, because the Gain and Offset
values are not limited in this case and incorrect Gain or Offset values may be inserted.
Upload button
– Press this button to upload Gain and Offset values from the target
platform. All 16 analog inputs gain and offset values will be uploaded.
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...