MAN205-0195042EN Rev. A
Aug. 2011
Page 53 of 137 pages
Heat T.O.
– Maximum time for heating the chamber for starting sterilization
When the duration of the heating stage is longer than Heat T.O., the program will stop
and display
the message ‗Low Heat‘.
Entry code
11
Resolution
1 minute
Minimum value
20 minutes
Maximum value
60 minutes
Ex Mode
– The method for exhausting the steam at the end of the process
Entry code
11
Ex Mode
Fast Ex
1
N.A.
2
Slow Ex. (waste)
3
Slow Ex. (liquids)
4
Cooling with compressed air
5
Cooling with water circulation
6
EndTemp
– Ending Temperature (Exhaust Stage)
Defines the temperature at which the process ends for opening the door.
Entry code
11
Resolution
1º
Minimum value
40ºC (104ºF)
Maximum value
137ºC (278ºF)
PulsNum.
– No. of pulses in the pre-vacuum stage
Entry code
11
Resolution
1
Minimum value
1
Maximum value
10
VacDip1
– Vacuum value in the first pulse
Defines the vacuum value in pulse No.1 of the pre-vacuum stage
Entry code
11
Resolution
1 kPa (0.1 psi)
Minimum value
1 kPa (0.15 psia)
Maximum value
75 kPa (10.8 psia)
Содержание 6671130
Страница 2: ...Web site http www tuttnauer com...
Страница 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Страница 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Страница 43: ...MAN205 0195042EN Rev A Aug 2011 Page 37 of 137 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Страница 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Страница 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...