Page 43 of 117 pages
4.1.2 Display
The CAT2007 display consists of 4 alphanumeric rows with 20 characters in each row:
First Row
First 3 characters define the Cycle no. (ranges from 1-14)
There are 8 characters to define the Cycle name.
The final 5 characters at the right of the row are allocated to displaying the status
of the appliance or the stage in which the cycle is found whilst running.
♦
When the autoclave is in waiting stage – ST.BY
♦
When the autoclave is ready to work – READY
♦
When a process ends correctly – END
♦
When a program ends incorrectly – FAIL
The names of the stages follow:
♦
Vacuum
♦
Exh. (Exhaust)
♦
Heat
♦
Dry
♦
Ster. (sterilization)
♦
Test
Following are the 14 cycles, names and relevant temperatures:
Cycle no.
Program
Temp. ºF
01-
Unwrapped
250
02-
Unwrapped
250
03 &4
Unwrapped
270/274
05-
Wrapped
270
06-
Wrapped
270
07-
Wrapped
274
08-
Wrapped
270
09-
Liquids
250
09-
Liquids
250
10-
Liquids
250
11 & 12-
Liquids
250
13
B&D test
14-
Leak test
Second Row:
This row displays the temperature in the chamber, pressure in the chamber, and
jacket pressure.
Third Row:
This row is used to display messages to the operator such as DOOR1 OPEN.
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...