Page 84 of 117 pages
Symptoms
Possible causes
check-ups and tests
Corrections
11. Drying incomplete
packs remain wet.
11.1 Insufficient steam pressure
to jacket.
11.1 Set or fix the
pressure transducer.
11.2 Insufficient vacuum.
11.2 See problems related
to vacuum par. 2 and
4.
11.3 Steam trap does not
eliminate the condensate.
11.3 Clean or repair the
steam trap.
11.4 Incorrect loading of the
autoclave.
11.4 Reload the autoclave
appropriately.
12. Problems related to
the control system.
12.1 Check if the flat cable or
MOLEX connectors
between the control boards
and keyboard are well
plugged in.
12.1 Plug in firmly the
MOLEX connectors.
12.2 Power supplies out of order.
12.2 Fix or replace the
power supply.
12.3 Control boards or keyboard.
12.3 Replace defective
board of
microcomputer unit.
12.4 Paper to printer missing or
quality unsuitable.
12.4 Insert a paper roll
inside the printer.
Ensure the
appropriate type of
paper is used.
12.5 Flat cable connection
between control board and
printer is not plugged in
correctly.
12.5 Plug in connection or
replace it.
12.6 Printer damaged.
12.6 Replace the printer.
12.7 Temperature sensor
disconnected or loose
connection. Sensor
damaged.
12.7 Fix connection or
replace sensor.
12.8 Pressure transducer, bad.
Mechanical or electrical
connections or transducer
damaged.
12.8 Fix connection or
replace transducer.
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...