11.2 Safety Tests after Repair
ATTENTION!
After every repair or dismantling the enclosure, the autoclave should pass
two safety electrical test by the Service Engineer. The following shall be
performed:
1. Enclosure Leakage Current Test.
Every autoclave should pass this test. The electrical potential of the
testing instrument should be 500V and the test should be performed
using the Megger The insulation resistance should be at least 2 M
Ω
.
The test is successful if there was no leakage.
2. Protective Earth Impedance Test
In the test should be measured the resistance between grounding
screw on the rear plate (or any other metallic part) and the grounding
pin of the power cord plug using the Ohmeter. The resistance should
not exceed 0.3
Ω
.
After performing these tests, the Service Engineer should complete
and sign the Work Order.