Page 89 of 117 pages
11.7 Cleaning and Replacing the Water Level Electrodes
The two electrodes of the water level control system are located on
top of the STS generator.
Cleaning
1.
Turn off the electric power and release the steam
pressure from the autoclave and the steam
generator.
2.
Disconnect the wire from the electrodes
connections.
3.
Open the nuts that tighten the electrodes.
4.
Insert a new electrode and tighten the fixing bolt
to avoid leakage.
5.
Reconnect the wires to the electrodes
Replacing
1.
Remove the two electrodes. Keep in mind the
place of the short and long electrodes.
2.
Clean the electrodes.
3.
Reassemble the two electrodes into their places.
4.
Tighten the nuts to avoid leakage.
5.
Reconnect the wire to the electrodes.
6.
Test the unit.