4.3.4 Changing Cycle Parameters
The screen that allows for parametric changes appears as follows:
Pressing Clear or Enter before entering the code returns to the Stand By screen,
and the RUN & FAIL LED's are lit.
Enter the correct code only via the UP/DN keypads, press ENTER to proceed to
the next screen.
Pressing the UP/DN keypads can also change all parameter values.
Cycle Setup Parameters:
Code 1 or 11 Parameters
STE Temp – Required sterilization temperature for the process
Entry code
1 or 11
Resolution
34ºF
Minimum value
221ºF
Maximum value
280ºF
STE Time – Required sterilization time for the process
Entry code
1 or 11
Resolution
1 minute
Minimum value
Ø minutes
Maximum value
59 minutes
Dry Time – Required drying time for the process:
Entry code
1 or 11
Resolution
1 minute
Minimum value
Ø minutes
Maximum value
59 minutes
Heat T.O. – Maximum time for heating the chamber for starting sterilization
When the duration of the heating stage is longer than Heat T.O., the program will
collapse with the message ‘Low Heat’.
Entry code
11
Resolution
10 seconds
Minimum value
1200 seconds (20 minutes)
Maximum value
3600 seconds (minutes)
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...