Page 67 of 117 pages
We are interested in correcting the measured temperature for 273
º
F without
changing the temperature at 250
º
F which is accurate.
Considering the above example, setting data can be done as follows:
Upper Row
A 140º F
R 141ºF
A 266º F
R 266ºF
The system calibrates the new offset and gain and loads them into the
non-erasable memory.
As a result, while running a 273
º
F program the correction is automatically
done, while at 250
º
F no change is felt.
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...