Page 35 of 117 pages
Program: 11 & 12 - Slow 250
Front door close Msw
PT100 Drain Control 5
PT100 Condense 1
PT100 Condense 2
IMT Jacket
IMT Chamber
Open door
Close door
Ejector
Buzzer
13 – Cool drain
15 – Water to ejector
43 – Air inlet
73 – Fast exh
74 – Slow exh.
75 – Vacuum
97 – Steam to jacket
93 – Steam to chamber
273ºF / 310kPa
250ºF / ~ 210kPa
199 kPa
221ºF / ~ 115kPa
100kPa
50kPa
25kPa
0kPa
Start (Close door)
Door secure
Temp. control
Pres. control
On (activated)
250ºF 15min.
Condition to open
T<230ºF ; P<115kPa
Ex Mode 1
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...