4.3.2 Selecting a Cycle
A cycle can only be selected during Stand By from the Stand By / Ready screen
which displays the cycle number and name in the top row.
A cycle can be selected as follows:
♦
Press the UP/DN keypad to change the cycle where pressing UP moves to
the next program and pressing DOWN moves to the previous program.
♦
Press on a number. When pressing on keypad 4 the chosen program will be
number 4, and so on.
Selecting a 2 digit cycle:
When 2 numbers are pressed within 2 seconds of each other, then both the
numbers define the program number (when the program number is 10 or above).
For example; to chose program 12, number 1 must be pressed, and within 2
seconds number 2 must be pressed.
Show Parameters
After selecting the cycle, pressing the SHOW command key presents the following
screen:
where the message displayed in the 3rd row describes the following:
♦
the temperature; the sterilization temperature
♦
the time; the sterilization time
♦
D-Time; drying time.
This screen is displayed for 10 seconds after which it automatically returns to the
ST.BY screen.
This command key is not active for cycle 14 – Leak Test.
Once the following conditions have been fulfilled,
♦
There is enough steam pressure in the jacket.
♦
Enough air pressure for the pneumatic valves (option).
♦
Doors closed.
♦
Enough water pressure for cooling (option).
♦
Enough water in the water reservoir.
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...