Page 46 of 117 pages
25 – Vac Res. Empty
If during the prevacuum stage, drying stage, vac. test or leak
test it is revealed that the lower float switch in the vacuum
water reservoir is open for more than 1 second (so there is no
water entering the water reservoir and the pump has
therefore been closed to protect it).i.e. Vfloat –L = “1”
27 – First Close Door
This message appears if there was an attempt to perform
IN/OUT Test whilst the door was open.
190 – Water Too High
This message appears when the level of water in the
chamber is too high. The chamber float switch detects the
water level and in case it’s too high closes all the valves and
the FAIL indicator lights.
Informative Messages
Calibration
This message appears automatically on panel 2 during
calibration.
Closing Door Wait
Whilst an automatic door is closing, this message is displayed
in order that the operator will wait until the closing operation
of the automatic door ends.
Open Door Wait
Whilst an automatic door is opening, this message is
displayed in order that the operator will wait until the opening
operation of the automatic door ends.
Door1 Open
This message appears when Door1 is open. It will also
appear if the door is supposedly closed but there is not
enough pressure in the gasket to seal it properly.
Under Test
This message appears on panel 2 during the IN/OUT Test
and during SetUp.
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...