Page 45 of 117 pages
4.2
Description of Displayed Messages and Safety Measures
03 – Manual Stop
Message is displayed and the FAIL indicator lights after the
STOP key is pressed for more than 1 second during the cycle
(excluding the drying stage).
100 – Man.Stop
This message is displayed and the FAIL indicator lights after
the STOP key is pressed for longer than one second in the
drying stage.
05-Door Unlock
Message is displayed if either of the following occurs:
The START button is pressed and the door is not closed;
If during a cycle for eg. the pressure drops in the gasket.
06-Temp. Error
If the temperature sensor disconnects for any reason during
the heating stage.
Indications that there is a disconnection are if the read
temperature is either higher than 140ºC or lower than 5ºC.
07- Low Heat
Message is displayed and sterilization does not start if the
autoclave has not reached sterilization temperature after
heating for the maximum time as defined in the HEAT T.O.
parameter.
08- Low Temp.
Message is displayed, fail indicator lights and cycle is
aborted, if the temperature drops for more than 5 secs. below
the required sterilization temperature.
09- High Temp.
Message is displayed, fail indicator lights and program is
aborted: If the temperature rises 37.4ºF above the required
sterilization temperature during the sterilization cycle.
10- Low Pres.
Message is displayed, fail indicator lights, and the program is
aborted if the pressure drops for more than 5 secs. below the
pressure correlated to the required sterilization pressure.
11-High Pres.
This message is displayed, fail indicator lights, and the
program is aborted, if pressure rises above the pressure
correlated to the sterilization temperature plus +37.4ºF- for
more than 5 secs.
12-Low Vacuum
This message is displayed if after 18 mins. the system has
not reached VacDip1or VacDip2 values during the prevacuum
stage. It also appears if the Leak Test program fails.
13 – Jacket not Ready
This message is displayed if the START key was activated
before the jacket reached the desired pressure.
21 – Door1 Safe Switch
If whilst opening door1 of an automatic sliding door the door1
switch closes, the door will open and this message will
appear.
23 – Door1 Gasket Low
Vacuum
If whilst opening Door1, you do not achieve the proper
vacuum in Gasket1.
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...