Page 57 of 117 pages
Puls pres2 – The pressure in pulse 2 and onwards in the prevacuum stage
After the completion of pulse no.1, the system will enter steam by the defined
value in puls press.
Entry code
11
Resolution
1 kPa
Minimum value
100 kPa
Maximum value
200 kPa
Stay Time - Waiting time between heating and sterilization
This parameter allows partial entry into the sterilization stage with a reduction of
overshoot and the reduction difference between the different sensors that are
entered into the chamber for control or validation.
In the heating stage the system stops heating before entering sterilization
according to the STAY TIME value. When this value is 2 minutes, the system
stops at 35.6ºF before entering sterilization and will proceed at a rate of
33.80 ºF/min during the STAY TIME.
Entry code
11
Resolution
1 min.
Minimum value
Ø min.
Maximum value
10 min.
This is the final parameter under Code 11. After pressing ENTER the display will
return to the ST.BY screen
Dry Vac. -
Dry Vac. =0 ; Vacuum pump works continuously.
Dry Vac. = 5; Vacuum pump works for 5 mins. after which air is introduced for 30
secs. of ON- and 30 secs. OFF intervals (depending on the interval selected by
the operator).
Entry code
11
Resolution
1 min.
Minimum value
Ø
Maximum value
90
Code 13 Parameters:
- The addition to the required pressure for sterilization
This parameter allows the pressure point that will be maintained during the
sterilization stage to be set in the following manner:
If the desired temperature is 273ºF, the required pressure (from the tables) is
305kPa.
When the is Ø, the system will be maintained at 305kpa during the
sterilization stage. When the equals 5kPa, the system will be
maintained at 310kPa in the sterilization stage. It is advised to work with a value
of 7kPa.
Entry code
13
Resolution
1 kPa
Minimum value
Ø
Maximum value
20 kPa
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...