Page 56 of 117 pages
VacTime2 – Waiting time for the remaining pulses
In the remaining pulses (not pulse no.1) in the prevacuum stage, after reaching
VacDip2 stage there is a time delay known as VacTime2 (in seconds).
Entry code
11
Resolution
1 second
Minimum value
1 second
Maximum value
1800 seconds (30 minutes)
Wtr Time – Time for entering the water into the generator after the autoclave
is touching the water
This value defines in seconds the time that the water pump continues working and
entering water to the steam generator after the upper electrode touches water,
to prevent a situation wherein immediately after the electrode touches water the
pump immediately begins operation.
Entry code
11
Minimum value
Ø second
Maximum value
90 seconds
LimitP add – Adding pressure to the maximum pressure
This value defines addition to the maximum pressure that is computed to the
jacket control.
For example: When the desired temperature is 273ºF, the maximum temperature
maintained in the jacket will be adjusted to a temperature of 137ºC (i.e. 332kPa)
When this pressure is not enough, it can be increased by changing the value of
the LimitP add.
If the parameter value remains Ø, the maximum pressure in the jacket will be
332kPa, however if this value will be 10, the maximum pressure will be 342kPa.
Entry code
11
Resolution
1 kPa
Minimum value
ØkPa
Maximum value
30 kPa
Calib Code:
This parameter is used to set the calibration code.
Entry code
11
Resolution
1
Minimum value
0
Maximum value
255
Puls Pres1 – The pressure in pulse no.1 in the prevacuum stage
This defines the first pulse pressure in the prevacuum stage.
If it is defined as 100kPa, the system will not enter steam into the chamber,
however the vacuum pump will immediately operate to create a vacuum.
If the value is defined over 100kPa, the system will open the steam valve to the
chamber until the desired pressure is reached.
Entry code
11
Resolution
1kPa
Minimum value
100 kPa
Maximum value
200 kPa
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...