Page 55 of 117 pages
Ex Mode – The method for exhausting the steam at the end of the process
Ex Mode
Fast Ex
1
N.A.
2
Slow Ex. (waste)
3
Slow Ex. (liquids)
4
Cooling with compressed air
5
Cooling with water circulation
6
EndTemp – Ending Temperature (Exhaust Stage)
Defines the temperature at which the process ends for opening the door.
Entry code
11
Resolution
34ºF
Minimum value
104ºF
Maximum value
279ºF
PulsNum. – No. of pulses in the prevacuum stage
Entry code
11
Resolution
1
Minimum value
1
Maximum value
5
VacDip1 – Vacuum value in the first pulse
Defines the vacuum value in pulse no.1 of the prevacuum stage.
Entry code
11
Resolution
1 kPa
Minimum value
1 kPa
Maximum value
75 kPa
VacDip2 – Vacuum value in the remaining pulses
This value defines the vacuum in the remaining pulses in the prevacuum stage
(not just pulse no. 2)
Entry code
11
Resolution
1kPa
Minimum value
1kPa
Maximum value
75kPa
VacTime1 – Waiting time in the first pulse
In the prevacuum stage, in the first pulse after reaching VacDip1, there is a waiting
time known as VacTime1 (in seconds).
This delay occurs so vacuuming can continue.
Entry code
11
Resolution
1 second
Minimum value
3 seconds
Maximum value
1800 seconds (30 minutes)
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...