Error Opening Type file:
File Name
Description:
This message may be displayed during download process, when
the program is trying to open ADMC compatible type file (*.mdb) in order to get
parameters data.
Reason:
The file may be damaged or used by another application.
Tips:
Try to download another file. If operation succeeds, try to find what
application is using the file and in the worst case, restart the PC. If the problem
continues, probably the files are damaged. In this case try to replace the file with
an identical file from the ADMC installation CD under 'Support' directory.
Invalid Actual High value!
Invalid Actual Low value!
Invalid Read High value!
Invalid Read Low value!
Description:
This message is displayed when the user is trying to calculate Gain
and Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
Wrong number or non- numeric data has been inserted in one of the
text boxes.
Tips:
Type the right Numeric data in the text box.
The Actual High value is not in range!
The Actual Low value is not in range!
The Read High value is not in range!
The Read Low value is not in range!
Description:
This message is displayed when the user is trying to calculate Gain
and Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
A wrong number has been inserted in one of the text boxes. A right
number must be: 1<Number<1000.
Tips:
Type the right number that is included in the above range.
Read High must be equal or great than Read Low!
Description:
This message is displayed when the user is trying to calculate Gain
and Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
Read High value is lees than Read Low value
Tips:
Read High value
must be
greater than Read Low value.
Can not calculate. Invalid values found! …
Description:
This message is displayed when the user is trying to calculate Gain
and Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
Read High value - Read Low value = 0
Tips:
Read High value
must be
greater than Read Low value.
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...