Page 75 of 117 pages
10.
PRINTER
10.1 Printer Operation
The autoclave is equipped with a character printer, which prints
a detailed history of each cycle performed by the instrument (for
the record or for subsequent consideration).
The printing is made on thermal paper with 24 characters per
line and contains the following information:
♦
Software version (is printed only on Power Dn)
♦
Real time
♦
Selected program
♦
Sterilization pressure
♦
Sterilization temperature
♦
Sterilization time
♦
Summary of performed cycle and identification hints.
When the sterilization cycle begins the printer starts printing the
above data.
After the preliminary printing, the autoclave starts performing the
sequence of operations of the cycle. The measured values of
temperature and pressure are printed at fixed time intervals,
according to the phase of the process, as shown in the table
below.
The data is printed from the bottom up, beginning with the date
and ending with “O.K.” for a complete cycle or “FAIL” for an
aborted cycle.
For an example of a typical printout, see next page.
Содержание 2301006
Страница 2: ......
Страница 14: ...Page 9 of 117 pages...
Страница 45: ...Page 40 of 117 pages 4 CONTROL AND MONITORING Control Panel CAT 2007...
Страница 79: ...Page 74 of 117 pages Control Panel Assembly...
Страница 84: ...Page 79 of 117 pages SERVICE INSTRUCTIONS...
Страница 99: ...Page 94 of 117 pages 11 11 Dismantling the Cabinets Side Panels Push the side cover upward and pull out A...
Страница 105: ...VERTICAL SLIDING DOOR Page 100 of 117 pages...
Страница 106: ...Page 101 of 117 pages...
Страница 107: ...Page 102 of 117 pages...
Страница 109: ...Page 104 of 117 pages XPCS Manual...