II-28
e. Adjust measured value
e-1 Zero Shift
Display Menu No. : INP-151 : +ZERO SHIFT
: INP-154 : -ZERO SHIFT
Range of Input
: +ZERO SHIFT : -9999.0 to 9999.00
: -ZERO SHIFT
: -9999.0 to 9999.00
Flow Rate Unit
: According to Measuring Unit that was set in
INP-121 : MULTIPLIER &INP-122 : FLOW UNIT.
Note :
In case of input into “-LOW CUT”, the character at the left-end in the LCD can be a “blank” or
“minus”.
e-2 Span Correction
Display Menu No. : INP-152 : +SPAN CORR.
: INP-155 : -SPAN CORR.
Initial value
: “ 1.000 ”
Range of Input
: 0.100 to 2.000
Individually setup
both Forward and Backward
Individually setup
both Forward and Backward